| Sign In | Join Free | My lightneasy.org |
|
Brand Name : PR
Model Number : PR 5437A
Place of Origin : Denmark
MOQ : 1
Price : 100usd
Payment Terms : T/T
Supply Ability : 99pcs 4-6 weeks
Delivery Time : 5-7 work days
Display : Optional LCD
Net Weight : 7 oz
Actuator Type : Side Rotary - No Lever
Corrent : 4-20 mA
Overvoltage Protection : Output protected at 110% to 120%
Zero & Span Adjustment : Field adjustable
Relative Humidity : 5 to 95%, non-condensing
Measurement Range : -200°C to 850°C
Packing : New and origin with factory seal
Input Protection : Internally fused, non-replaceable
Ingress Protection : IP65/IP67
Maximum Allowable : 500 psi (35 bar)
Storage Temperature : -40 to 85°C (-40 to 185°F)
Origin : China
Hs Code : 85363000
| Operating temperature | -50°C to +85°C (standard) -40°C to +80°C (SIL) |
| Storage temperature | -50°C to +85°C |
| Calibration temperature | 23...25°C |
| Relative humidity | < 99% RH (non-condensing) |
| Protection degree (enclosure/terminal) | IP68 / IP00 |
| Dimensions | Ø 44 x 21.45 mm |
| Center hole diameter | Ø 6.35 mm / ¼ in |
| Weight approx. | 50 g |
| Wire size | 1 x 1.5 mm² stranded wire |
| Screw terminal torque | 0.4 Nm |
| Vibration | IEC 60068-2-6 2...25 Hz: ±1.6 mm 25...100 Hz: ±4 g |
| Supply voltage | 7.5*...48** VDC |
| Internal power dissipation | ≤ 850 mW |
| Additional min. supply voltage when using test terminals | 0.8 V |
| Min. load resistance at >37 V supply | (Vsupply - 37) / 23 mA |
| Isolation voltage, test / working | 2.5 kVAC / 55 VAC |
| Response time | 75 ms |
| Programmable damping | 0...60 s |
| Warm-up time | < 5 min. |
| Start-up time | < 2.75 s |
| Long-term stability | ±0.05% of span / year (±0.18% of span / 5 years) |
| Signal / noise ratio | > 60 dB |
| RTD type | Pt10...10000, Ni10...10000, Cu5...1000 |
| Cable resistance per wire | 50 Ω (max.) |
| Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
| Sensor current | < 0.15 mA |
| Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
| Cold junction compensation (CJC) | Constant, internal or external via a Pt100 or Ni100 sensor |
| Normal range, programmable | 3.8...20.5 / 20.5...3.8 mA |
| Extended range (output limits), programmable | 3.5...23 / 23...3.5 mA |
| Updating time | 10 ms |
| HART protocol revisions | HART 7 and HART 5 |
| ATEX | DEKRA 18ATEX0135X |
| IECEx | IECEx DEK. 16.0029X |
| CSA | CSA 16.70066266 |
| c FM us | FM16US0287X / FM16CA0146X |
| SIL | SIL 2 / SIL 3 certified & fully assessed acc. to IEC 61508 |
|
|
2 Wire HART 7 Temperature Transmitter PR 5437A With RTD TC Potentiometer Input IP65 / IP67 Images |